Leakage on mark test (electric mark the index test) is IEC60112, UL 746 A, GB/T4207, GB4706.1, ASTM D 3638-92, the standards of simulation experiment project.
Technical parameters:
1, electrode materials: brass electrode head-platinum.
2 and electrode pressure: 1.0 + /-0.05 N
3 and electrode distance:4.0 mm+ /-0.01 mm, Angle of 60 ° ± 5
4. The electrode voltage: 100 ~ 600 V (48 ~ 60 HZ) can be adjusted, short-circuit current in 1.0 + /-0.1 Avoltage down less than 8%
5, drops liquid height: for 30 to40 mmis adjustable
6, drops of liquid size: 44 ~ 55 drops/cm3 1 can be adjusted
7, drops of liquid time: 30 s ± 5 s,
8, drops of liquid number: 0 ~ 9999 times can be adjusted
9. Leakage judgment: A loop current more than0.5 Aand maintain 2 seconds the relay action to cut off the current, instructions, sample not qualified
10, test process: test procedures to be automatic control, independent convulsions
11, consult standard: IEC60112, GB/T4207, GB4706.1
12, studio volume: no less than 0.5 cubic
13, the equipment size:1100 mmwide x700 mmdeep x1300 mmhigh
14, power supply parameters: 1 KW / 220 VAC / 48-60 Hz
15 control: the single chip microcomputer + touch screen control, the high accuracy thermometer independent type electrical control system, strong anti-jamming capability.
16, operation efficiency: test control processes to achieve a key type operation, fast and efficient test
17, test standard integration: according to actual test process writing test procedures
TEL:15014035973