30032A钳型表
联系人:吴先生(15113461516)
电话:020-82348737\ 13710725812
网址:
日本横河YOKOGAWA授权一级代理!大量现货优势供应
30031A and 30032A Leakage Clamp-on Testers |
30031A钳型表30031A钳型表 |
|
|
|
· Can measure leakage AC currents of 1mA
· Can measure a broad range of currents from leakage currents, at a resolution of 0.001A, to load currents up to 60A.
· Digital display is enlarged by 20% compared to the previous models.
· Filter function can cut off the harmonics-currents components from 2nd order (for 30032A only) |
|
Specifications |
30031A钳型表
● AC current measurement
Range |
Resolution |
Accuracy |
Max allowable current |
3mA |
0.001mA |
0.010<1<32.70mA: |
3.270mA |
30mA |
0.01mA |
1.0% + 5 |
32.70mA |
30A |
0.01A |
0.05<1<50.0A: 1.0% + 5 |
32.70A |
60A |
0.1A |
50.0<1<60.6A: 5.0% + 5 |
60.6A |
● AC current measurement (for 30032A only)
Range |
Resolution |
Accuracy |
Max allowable current |
3mA |
0.001mA |
0.010<1<32.70mA: |
3.270mA |
30mA |
0.01mA |
1.5% + 5 |
32.70mA |
30A |
0.01A |
0.05<1<50.0A: 1.5% + 5 |
32.70A |
60A |
0.1A |
50.0<1<60.6A: 5.5% + 5 |
60.6A |
● Zero correction: 3mA range: displays 0.000mA(zero) when 0.010mA <1 30A range: displays 0.00A(zero) when 0.05A <1
● Filter specificatons (for 30032A only) Amplitude ratio at 100Hz: -38 dB (1.26%) or less (typical: -41 dB) Amplitude ratio at 120Hz: -53 dB (0.22%) or less (typical: -56 dB) |
General Specifications |
|
|
|
|
Parameter |
Specification |
Method of detection |
Mean value detection and rms-value calibration |
Display |
LCD |
Digital display: 3200 counts |
Bar-graph display: 32 segments |
Range Switching |
Range selection Auto or Manual |
Data hold |
On all ranges |
Operating temperature and humidity |
0 to 50 deg.C, 80% RH or less (no condensation) |
Temperature Coefficient |
Following values must be added in the temperature range of either 0 to 18℃ or 28 to 50℃. 0<1<50.0A: ±(0.08% of reading /℃ + 0.5 digits/℃) 50.0<1<60.6A:±(0.3% of reading/℃ + 0.5 digits/℃) |
Effect of external magnetic field |
0.0005% typical value (in terms of magnitude of current in adjacent wires) |
Safety Standard |
Conforms to EN 61010-1, EN 61010-2-032 CAT.III 300V Circuit voltage: 300 V rms or less |
Circuit Voltage |
300Vrms or less |
Withstanding Voltage |
3.7kV AC for one minute |
Power Supply |
One CR2032 (Coin-shaped) 3 V battery |
Power consumption |
6 mW max |
Battery Life |
Approx. 90 hours |
Automatic power-off |
power approx. 10 minutes after the last switch operation |
Diameter of measurable conductor |
40 mm at maximum |
Dimensions |
Approx. 70(W) × 179(H) × 25(D) excluding protrusions |
Weight |
Approx. 200g |
Accessories |
User's manual, battery, soft carrying case (RB057) | |