Output Current Range 0 to ±15 mA DC or peak AC. See Automatic Power Limit feature for limitations.
PERFORMANCE
DC Voltage Gain 4000 V/V.
DC Voltage Gain uracy Better than 0.1% of full scale.
Offset Voltage Less than ±4 V.
Output Noise Less than 5 V rms (measured using the true rms feature of the Hewlett Packard Model 34401A digital multimeter).
Slew Rate (10% to 90%, typical) Greater than 350 V/µs.
Large Signal Bandwidth (2% Distortion) DC to greater than 1.4 kHz.
Small Signal Bandwidth (-3dB) DC to greater than 20 kHz.
Stability
Drift with Time Less than 50 ppm/hr, noncumulative.
Drift with Temperature Less than 100 ppm/°C.
Settling Time (to 1%) Less than 200 µs for a 0 to 40 kV step.
Automatic Power Limit Automatically limits the internal power dissipation to protect the Model 40/15 from overheating. The following graph illustrates the Model 40/15 automatic output power limit range.