| 7130 | 7132 | 7140 | 7142 | |
| 5KVAC/20mA | ||||
| Range | Resolution | uracy | ||
| 0-5.00 | 0.01 | ±(2% of setting + 5V) | ||
| 50Hz/60Hz ±100ppm, User Selection | ||||
| Sine Wave ,THD.<2% (Resistive Load), Crest Factor = 1.3 - 1.5 | ||||
| ± (1% of output + 5V), From no load to full load | ||||
| 0 - 20.00 | 0.01 | ±(2% of setting + 2counts) | ||
| 0 - 9.999 | 0.001 | ±(2% of setting + 2counts) | ||
| 0.1 - 999.9 | 0.1 | ±(0.1% + 0.05sec) | ||
| 0 - 999.9 | 0.1 | |||
| 0, 0.3 - 999.9 (0=continuous) | 0.1 | |||
| 0, 1 - 9 ranges (0=OFF, 9 is the most sensitivity) | ||||
| 6KVDC/7500µA | ||||
| 0 - 6.00 | 0.01 | ±(2% of setting + 5V) | ||
| < 5%(6KV/7500µA at Resistive Load) | ||||
| 0 - 7500 | 1 | ±(2% of setting + 2counts) | ||
| 0 - 999.9 | 0.1 | ±(2% of setting + 2counts) | ||
| 0.1 - 999.9 | 0.1 | ±(0.1% + 0.05sec) | ||
| 0, 1.0 - 999.9 | 0.1 | |||
| 0, 0.4 - 999.9(0=continuous) | 0.1 | |||
| 0, 1 - 9 ranges (0=OFF, 9 is the most sensitivity) | ||||
| ≦ 200 msec | ||||
| 1KVDC/9999MΩ | ||||
| 30 - 1000 | 10 | ±(2% of setting + 5V) | ||
| 0, 1 - 9999 (0=OFF) | 1 | ±(2% of setting+2counts)at 500-1000VDC,1 -1000 MΩ | ||
| 1 - 9999 | 1 | ±(5% of setting+2counts)at 500-1000VDC,1000 -9999 MΩ | ||
| ±(8% of setting+2counts)at 30-500VDC, 1 -1000 MΩ | ||||
| 0.1 - 999.9 | 0.1 | ±(0.1% + 0.05sec) | ||
| 0, 1.0 - 999.9 | 0.1 | |||
| 0, 0.5 - 999.9(0=continuous) | 0.1 | |||
| 0 - 6.00 | 0.01 | ±(1.5% of reading + 1count) | ||
| 30 - 1000 | 1 | ±(1.5% of reading + 5V) | ||
| 0 - 3.500 | 0.001 | ±(2% of reading + 2counts) | ||
| 3.00 - 20.00 | 0.01 | |||
| 0 - 350.0 | 0.1 | ±(2% of reading + 2counts) | ||
| 0.300 - 3.500 | 0.001 | |||
| 3.00 - 7.50 | 0.01 | |||
| 1 - 9999 (Auto Range) | 0.001 0.01 0.1 1 | ±(2% of setting+2counts)at 500-1000VDC, 1 -1000 MΩ ±(5% of setting+2counts)at 500-1000VDC, 1000 -9999 MΩ ±(8% of setting+2counts)at 30-500VDC, 1 -1000 MΩ | ||
| 115/230Vac±15%, 50/60Hz ± 5%, Fuse 3.15A | ||||
| Input : Test , Reset , Memory1.2.3 , Interlock | ||||
| Output : Pass , Fail , Processing , Reset-Out | ||||
| RS485 Interface | ||||
| 10 memories, 3 steps/memory | ||||
| 128×64 Graphic LCD | ||||
| Built-in Smart GFI circuit,GFI trip current 450?A max, HV shut down speed: <1mS | ||||
| To prevent unauthorized alteration of the test Parameters | ||||
| Build-in software and external calibrated meters | ||||
| Build-in software driven verification menu to test fault detection circuits | ||||
| Range: 0-9 ;0=OFF, 1 is softest volume, 9 is loudest volume. | ||||
| 0-40℃, 20-80%RH | ||||
| 215mm(W)×89mm(H)×370mm(D)/ 10Kg | ||||
| ×1 | ||||
| ×2 (Including a spare contained in the fuse holder) | ||||
| ×1 | ||||
| ×1 | ||||
| ×1 | ||||
| *product specifications are subject to change without notice. | ||||