德国Analytikjena(耶拿) Biometra TB2干式加热器是德国耶拿公司生产的一款安全可靠的干式加热器。现在热卖中,如需购买,可联系我们!
产品简介:
·Dry block heater for incubation at constant temperature.
·Dual block positions, removable blocks
·11 different block options
·Temperature range: RT +5℃ to 130℃
·The Thermoblock TB2 has been developed for incubation of reaction tubes at fixed temperatures. ·Temperature control is achieved by means of a Pt1000 microsensor in order to reach outstanding block uniformity (+/- 0.1℃ at 37℃).
·Incubation of a magnitude of samples contained in different reaction tubes causes no problems for the block heater.
·The TB2 has a capacity for two single blocks or one combi block.
·Combinations of different single blocks allow simultaneous incubation of e.g. 1.5 ml and 0.5 ml reaction tubes.
·The waterproof front panel is slightly angled to facilitate data entry and temperature readings.
·It features 3 buttons and a central control knob for starting the block, entering temperatures up to 130℃ and additional functions.
·The large LED displays the current temperature as precise as up to one decimal.
·The TB2 has a small footprint of 200 x 280 x 100 mm (W x D x H).
Temperature range | 室温+5℃~130℃ |
Temperature setting | 15°℃~130℃ |
Temperature stability | +/- 0.1℃ (@ 37℃) |
Uniformity within the block | +/- 0.1℃ (@ 37℃) |
Uniformity across similar blocks | +/- 0.2℃ (@ 37℃) |
Heat up time | 30 - 37℃ | approx. 2 min, then allow 3 min for stabilisation |
| 25 - 100℃ | approx. 15 min, then allow 5 min for stabilisation |
| 30 - 130℃ | approx. 20 min, then allow 5 min for stabilisation |
Capacity | 2 removable aluminium blocks for test tubes or microcentrifuge tubes (Single blocks)or1 removable microtiter plate block (Combi block) |
Display | digital, LED |
Display resolution | 0.1℃ |
Control LEDs for | heater, fault |
Timer | 1 - 9,999 min |
Timer signal | buzzer |
External temperature sensor | option |
Additional functions | temperature deviation alarm, offset adjustment, delay start and stop, calibration |
Overtemperature protection | Internal thermal fuse |
Dimensions (W x D x H) | 200 mm x 280 mm x 100 mm |
Enviroment operating range | 10℃ to 35℃ (80 % max. relative humidity) |
Power | 220 - 240V/50Hz,300W |
Block Dimensions (W x D x H) | Single block: 100 x 70 x 62.5 mm Combi block: 140 x 100 x 49 mm |