1.4.产品图片Picture of EXD 1800B
EDX 1800B model机型
EDX 1800B Model
1.5.工作条件Working Condition
●工作温度:15-30℃ Working temperature: 15-30℃
●相对湿度:≤70% Relative humidity: ≤70%
●电 源:AC: 220±5V Power supply: AC 220±5V
1.6.技术性能及指标:Technical Performance and Specifications
1.6.1.元素分析范围从硫S到铀U(77种元素)
The elements can be analyzed from the sulphur (S) to uranium (U)
1.6.2.元素含量分析范围为1ppm到99.99%
The element content range from 1 ppm to 99.99% can be analyzed.
1.6.3.测量时间:60-200秒 Measurement time: 60-200s
1.6.4. RoHS指令规定的有害元素(限Cd/Pb/Cr/Hg/Br)其检测限度最高达2ppm;
Detecting for hazardouselements(onlyCd/Pb/Cr/Hg/Br)restricted in RoHS directive, the limit can reach 2ppm
1.6.5玩具重金属:铅(Pb)、汞(Hg)、镉(Cd)、铬(Cr)、溴(Br).砷(AS).,锑(Sb).硒(Se)钡(Ba)等
1.6.6.多次测量重复性可达0.1%(总荧光强度)(当样品含量大于96%);
The measurement repeatability on many occasions can reach 0.1% of total intensity of fluorescence when the content of samples is greater than 96%.
1.6.7.长期工作稳定性为0.1%;(总荧光强度)
Long-term working stability being 0.1% (total intensity offluorescence)
1.6.8.能量分辨率为135±5电子伏特Energy resolution: 135±5eV
1.6.9.温度适应范围为15℃至30℃ Suitable temperature range is from 15℃ to 30℃
1.6.10.电源:交流220±5V;(建议配置交流净化稳压电源。)
Power supply: AC 220±5V. The purified AC regulated power supply is recommended.)
1.6.11.相互独立的基体效应校正模型 Independent matrix effect correction model
1.6.12.多变量非线性回归程序Multi-variable non-linear regression procedure
1.6.13.任意多个可选择的分析和识别模型 Arbitrary optional analysis and identification models
1.6.14.一次可同时分析24个元素 Ability toanalyze24 elements simultaneously
1.6.15仪器误差:在测试含量在100ppm内的时候误差会在正负10ppm以内,含量在100-1000ppm以上时误差会在正负10%以内,,含量在1000ppm以上时误差会在正负15%以内
2.仪器硬件部分主要配置Main Hardware Configuration of EDX 1800B
2.1 硅针半导体探测器+放大电路:(最新型的探测器)
Silicon Pin Semiconductor Detector + Amplifier Circuit(streamlined detector)
技术指标及功能: Technical Specifications and Features:
2.1.1.电制冷半导体探测器;分辨率:135±5电子伏特Electro-refrigeration semiconductor detector; Resolution: 135±5eV
2.1.2.对样品特征X射线进行探测;把探测采集的信息进一步放大。对样品的检出限、测试精度大大提高。Detecting characteristic X rays of the samples, and amplifying the information from the detection. The detection limit is reduced and detection accuracy is increased greatly.
2.2 X光管X-Ray Tube
技术指标及功能: Technical Specifications and Features:
2.2.1.使用寿命大于2万小时Service life is longer than 20,000 hours.
2.2.2.产生X射线激发样品Emitting X rays to excite the samples
2.3高、低压电源High/Low Voltage Power Supply
技术指标及功能: Technical Specifications and Features:
2.3.1.从5~50kV,1毫安 5~50kV, 1mA
2.3.2.为X光提供专用电源 Providing power supply exclusively to X-ray