供应其他PS300-12氙灯专用电源PS300-12C
| Input voltage | 90 – 264V rms. universal input |
| AC Power Connector | 3 X 0.25” spade tabs |
| Input Frequency | 47 – 63Hz |
| Maximum Input Power | <500W |
| Maximum Input Current | <7A rms. |
| Input Fuse | 8A time lag (T8A 250V HRC 5 X 20mm in line only) |
| Input Power Factor | >0.95 (meets the harmonic requirements of EN61000-3-2 class-D) |
| Mains Earth Leakage Current | <120uA @120Vrms, <240uA @240Vrms |
| Maximum Output Voltage | 22Vdc (compensates automatically for lamp voltage up 22V max) |
| Output Current Range | 4 – 23A (constant current, adjustable by internal pot or remotely) |
| Output Connections | 2 X 0.25” spade tabs |
| Over Current Protection | >25A constant current |
| Maximum Output Power | 320W lamp + 30W auxiliary / fan |
| Output Ripple Current | <2% peak –peak |
| Lamp Boost Voltage | >200Vdc |
| Lamp Ignition Pulse | Bipolar >+/-15KV wrt ground (30KV peak – peak) |
| Interface Controls | Remote Enable / Disable. Auto / Manual mode. Interlock Remote output current adjust. Current monitor |
| Interface Signals | LAMP ON and LAMP REPLACE signals |
| Auxiliary Outputs | 5V +/-4% @ 1A, 12V+/-4% @ 2A isolated & independent of lamp operation. |
| Fan Output | 12V +/-4% @ 1A (only present when lamp operating) |
| Efficiency | 70% (at full power) |
| Operating Temperature Range | 0 – 45deg.C |
| Storage Temperature | -20 to +85deg.C |
| Over Temperature Protection | Thermostat shuts psu down if excessive internal temperature. |
| Air Flow Requirement | >60cfm |
| Maximum Operating Humidity | 95% RH non condensing |
| Shipping Shock & Vibration | ISTA1A, 2A |
| Safety Compliance | UL60601-1, EN60601-1, IEC60601-1, CSA-22.2 No.601-1-M90 CE mark, CB report. (psu requires correct installation) UL file. E193746 |
| Size | 7” X 4.5” X 2.05” (177.8 X 114.2 X 52mm) |