Detail |
General | ReactionResponse Network | Let-GoResponse Network | Unweighted Burn Hazard
Response Network |
Equivalent network component values: | Complies with values
specified in IEC 990 (excluding fuse) |
Equivalent measuring instrument load: | 1 M_ - 10pF | 1 M_ - 62 pF | 1 M_ -
1 pF |
Ranges | 0.3, 1, 3,10 M.I.U. | 0.3, 1, 3,10 M.I.U. | 0-100 mA RMS |
Current accuracy: | ±2% F.S. @ 60
Hz |
Meter measurement method: | True RMS |
Meter frequency response: | (Relative to ANSI C101-1992 or UL-1459 2nd edition) |
DC to 1 Hz | Pointer shall track within 5% of peak |
2Hz to 19Hz (Accuracy not supported) | |
20 Hz to 200 KHz | ±2% F.S. | ±2.5% of F.S. | ±2% F.S. |
200 KHz to 1 MHz | ±2% F.S. | ±2.5% F.S | ±5% F.S. |
Output sensitivity: | Full scale meter indication equals 1V RMS (Measured
with a1 M_,
12 pF load) |
Output accuracy: | ±2% of reading
@ 60 Hz |
Output frequency response: | (Relative to ANSI C101-1992 or UL-1459 2nd edition) |
DC to 50Hz | ±2% of reading |
50 Hz to 100 KHz | ±2.5% of reading | ±2% /-3% of reading | ±2% of reading |
100 kHz to 1 MHz | ±5% of reading |
Voltmeter
range: | 0-300 V (AC or DC) |
Voltmeter accuracy: DC
to 1 Hz | Pointer shall track within 5% of peak |
2Hz to 19Hz (Accuracy not supported) | |
20 Hz to 1KHz _ | 3% F.S. @ 60 Hz (Add _1% for every
additional 100Hz) |
Voltmeter frequency response: | DC to 1 KHz |
Power requirements | (2) 9V (NEDA1604A)
batteries |
Physical | 7" x 5.25" x 3.125", 2-1/2 lbs(1.4kg), ABS plastic, fused input |
Environmental | 27°C _2° C, 70% non condensing relative humidity. Outside
this range derate by 10% per °C |
Maximum
operating range | 0°
to 40° C |